ALLERGAN (COSTA RICA) INSPIRA SMOOTH COHESIVE SILICONE GEL FILLED BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
|
Back to Search Results |
|
| Catalog Number |
SCF-240 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Pain (1994)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.The event of pain is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Reason for reoperation due to "pain, hardness, and swelling." further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Allergan did not submit this mdr within 30 days of becoming aware.Recent stimulated reporting related to 2011068-7/2/19-001-r has increased complaint and mdr volume.Allergan is implementing a plan to address the increased volumes.
|
| |
|
Event or Problem Description
|
|
Patient reported right side "pain, hardness, and swelling." device has been explanted.
|
| |
|
Search Alerts/Recalls
|
|
|