DEPUY IRELAND - 9616671 ATTUNE RP TIB BASE SZ 9 CEM; ATTUNE IMPLANT : KNEE TIBIAL TRAY Back to Search Results
Model Number 1506-10-009
Medical Device Problem Code Loss of or Failure to Bond (1068)
Health Effect - Clinical Codes Pain (1994); Limited Mobility Of The Implanted Joint (2671); No Code Available (3191)
Date of Event 10/24/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
On (b)(6) 2017, the patient underwent total left knee arthroplasty due to degenerative joint disease and valgus.The patella was resurfaced.The surgeon reported no intraoperative complications.The patient was implanted with the attune knee system and smartset bone cement x 2.On (b)(6) 2019, the patient underwent a left knee revision due to loosening of the tibial component, lack of motion, fusions, and pain.The surgeon reported the tibia had bone that had grown up around the tibial tray as if the body was trying to stabilize a loose implant.The tibial tray was easily removed and had no cement adherent at the tray to cement interface.The cement was well-fixed to the bone.The femoral component was removed with minimal bone loss.The surgeon indicated the patella was revised.The patient was implanted with depuy revision knee system.Competitor¿s cement was utilized, and there were no complications reported.Doi: (b)(6) 2017.Dor: (b)(6) 2019, (lt knee).
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.Depuy synthes considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE RP TIB BASE SZ 9 CEM
Common Device Name
ATTUNE IMPLANT : KNEE TIBIAL TRAY
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
MDR Report Key9496793
Report Number1818910-2019-124103
Device Sequence Number19718296
Product Code NJL
UDI-Device Identifier10603295042631
UDI-Public10603295042631
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
P83005
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2019
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/30/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1506-10-009
Device Catalogue Number150610009
Device Lot Number8418666
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 12/03/2019
Supplement Date Received by Manufacturer09/01/2020
Initial Report FDA Received Date12/19/2019
Supplement Report FDA Received Date09/02/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age54 YR
-
-