• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDOS INTERNATIONAL SàRL TRUESPAN 12 DEGREE PLGA; SOFT-TISSUE ANCHOR, BIOABSORBABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SàRL TRUESPAN 12 DEGREE PLGA; SOFT-TISSUE ANCHOR, BIOABSORBABLE Back to Search Results
Model Number 228161
Device Problem Activation, Positioning or Separation Problem (2906)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/04/2019
Event Type  malfunction  
Manufacturer Narrative
Device was used for treatment, not diagnosis.Udi: (b)(4).The expiration date is not currently available.As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported by the sales rep that during a meniscal repair procedure, it was observed that the truespan 12 degree device misfired.There was no delay in the surgical procedure.There was a spare device available for use to complete the surgery.There was patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: investigation summary
=
> the complaint device is not being returned, it was discarded by the customer, therefore unavailable for a physical evaluation.This complaint cannot be confirmed.With the information provided, and without the complaint device to evaluate, we cannot determine a root cause for the reported failure.A manufacturing record evaluation was performed for the finished device lot number (4l51637), and no non-conformances were identified.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.Device history lot
=
>a manufacturing record evaluation was performed for the finished device lot number (4l51637), and no non-conformances were identified related to this complaint.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRUESPAN 12 DEGREE PLGA
Type of Device
SOFT-TISSUE ANCHOR, BIOABSORBABLE
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL
chemin-blanc 38
le locle 02400
SZ  02400
MDR Report Key9497340
MDR Text Key205271834
Report Number1221934-2019-60064
Device Sequence Number1
Product Code MBI
UDI-Device Identifier10886705026043
UDI-Public10886705026043
Combination Product (y/n)N
PMA/PMN Number
K153667
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 12/04/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/19/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number228161
Device Catalogue Number228161
Device Lot Number4L51637
Was Device Available for Evaluation? No
Date Manufacturer Received01/22/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-