• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. ULTRASCORE FOCUSED FORCE PTA BALLOON

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. ULTRASCORE FOCUSED FORCE PTA BALLOON Back to Search Results
Model Number US35130630
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Occlusion (1984); Vascular Dissection (3160)
Date of Event 11/26/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
As the lot number for the device was provided, a manufacturing review will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.(expiry date: 10/2020).
 
Event or Problem Description
It was reported through the results of a clinical trial, that during an angioplasty procedure with a focused force percutaneous transluminal angioplasty (pta) balloon dilatation catheter in the right femoral artery, type a dissection was identified.It was further reported that intervention was performed.Patient was discharged the same day as the procedure.
 
Additional Manufacturer Narrative
H10: manufacturing review: the lot number was provided and a lot history review was conducted.Investigation summary: the device was not returned for evaluation.Images and medical records were not provided.The investigation is inconclusive, as no objective evidence has been provided to confirm any alleged deficiency with the device.Based upon the available information, the definitive root cause is unknown.Labeling review: the review of the ifu (instructions for use), indications, warnings, precautions, cautions, possible complications, and contraindications showed that the product labeling is adequate.H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported through the results of a clinical trial, that during an angioplasty procedure with a focused force percutaneous transluminal angioplasty (pta) balloon dilatation catheter in the right femoral artery, type a dissection was identified.It was further reported that intervention was performed.Patient was discharged the same day as the procedure.Approximately six months post index procedure, duplex ultrasound demonstrated target lesion occlusion.The current patient status was not provided.
 
Event or Problem Description
It was reported through the results of a clinical trial, that during an angioplasty procedure with a focused force percutaneous transluminal angioplasty (pta) balloon dilatation catheter in the right femoral artery, type a dissection was identified.It was further reported that intervention was performed.Patient was discharged the same day as the procedure.Approximately six months post index procedure, duplex ultrasound demonstrated target lesion occlusion.The current patient status was not provided.
 
Additional Manufacturer Narrative
H10: manufacturing review: the lot number was provided and a lot history review was conducted.Investigation summary: the device was not returned for evaluation.Images and medical records were not provided.The investigation is inconclusive, as no objective evidence has been provided to confirm any alleged deficiency with the device.Based upon the available information, the definitive root cause is unknown.Labeling review: the review of the instructions for use, indications, warnings, precautions, cautions, possible complications, and contraindications showed that the product labeling is adequate.H10: (expiry date: 10/2020).H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRASCORE FOCUSED FORCE PTA BALLOON
Common Device Name
FOCUSED FORCE PTA BALLOON
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key9500431
Report Number2020394-2019-05702
Device Sequence Number1542016
Product Code PNO
UDI-Device Identifier00801741116926
UDI-Public(01)00801741116926
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K172571
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,study
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 01/07/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberUS35130630
Device Catalogue NumberUS35130630
Device Lot NumberGFCX2169
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 12/02/2019
Supplement Date Received by Manufacturer08/31/2020
12/17/2020
Initial Report FDA Received Date12/20/2019
Supplement Report FDA Received Date09/02/2020
01/07/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age57 YR
Patient Weight86
-
-