| Model Number |
DL-330-10-K |
| Medical Device Problem Codes |
Signal Artifact/Noise (1036); Break (1069)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
09/03/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The explanted lead was not returned to neuropace for analysis.However, review of the device ecog and lead impedance confirmed that the ecog data are suggestive of a potential lead break.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2019, intermittent artifact was identified through ecog review, suggestive of a possible lead break in the left depth lead connected to port 2.Further review of the lead impedance and ecog data was performed and was consistent with a lead break.During a routine rns neurostimulator replacement procedure, the lead was explanted and a new lead was implanted and connected to port 2.
|
| |
|