• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DAVIS & GECK CARIBE LTD POLYSORB; ENDOSCOPIC TISSUE APPROXIMATION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DAVIS & GECK CARIBE LTD POLYSORB; ENDOSCOPIC TISSUE APPROXIMATION DEVICE Back to Search Results
Model Number 170051
Medical Device Problem Codes Entrapment of Device (1212); Human-Device Interface Problem (2949)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 12/02/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during a laparoscopic gastric bypass, the surgeon experienced resistance when going through the tissue as the surgeon toggled the needle between the jaws.The needle would get stuck as if there was a barb on the needle shaft that was caught on the tissue.Doctor mentioned that 3 of 4 instruments felt gritty upon needle exchange between the jaws when toggling.Tissue thickness appeared normal.The fourth and final load felt smooth and went through with out resistance and the surgeon continued to use the handle to complete the case.There was no patient injury.
 
Additional Manufacturer Narrative
H3 evaluation summary: post market vigilance (pmv) led an evaluation of one device.The visual inspection of the returned product noted one needle with several inches of suture attached.The needle was inspected under microscope where no abnormalities were found.Proper swage marks and properly formed loading slots were observed.The returned needle was found to function properly and remained engaged in the test instrument throughout testing.No difficulty was experienced in loading, unloading or toggling the returned needle.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.The investigation concluded there were no abnormalities that would have caused or contributed to the reported condition.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POLYSORB
Common Device Name
ENDOSCOPIC TISSUE APPROXIMATION DEVICE
Manufacturer (Section D)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DO  0101
MDR Report Key9503705
Report Number9612501-2019-02490
Device Sequence Number2981117
Product Code OCW
UDI-Device Identifier20884521126753
UDI-Public20884521126753
Combination Product (Y/N)N
Initial Reporter StateHI
Initial Reporter CountryUS
PMA/510(K) Number
K934738
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/20/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/29/2024
Device Model Number170051
Device Catalogue Number170051
Device Lot NumberJ9C0257Y
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/23/2019
Initial Date Received by Manufacturer 12/09/2019
Supplement Date Received by Manufacturer02/06/2020
Initial Report FDA Received Date12/20/2019
Supplement Report FDA Received Date02/20/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-