| Model Number |
6532 |
| Medical Device Problem Codes |
Break (1069); Entrapment of Device (1212)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
12/11/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that catheter entrapment and shaft break occurred.A 15mm x 5.00mm nc quantum apex balloon catheter was selected for use.However, upon entering a 0.14 unknown guide wire into the catheter, the balloon shaft was stuck on the wire in the guide catheter and as the physician was pushing the device, the catheter broke.The physician was able to remove everything as he pulled back the wire.A new wire and a 20x5.00mm balloon was used and completed the procedure.No patient complications were reported and the patient's status was fine.
|
| |
|
Additional Manufacturer Narrative
|
|
Describe event or problem updated.Device evaluated by mfr.: returned product consisted of the distal end of a nc quantum apex balloon catheter and a portion of a filter guidewire.There was blood in the inflation lumen and balloon.The shaft was kinked and had numerous buckled and stretched locations throughout the shaft.The guidewire was unable to be removed due to the condition of the catheter shaft.The midshaft was separated, the end of the midshaft was jagged and stretched due to tensile overload.Measurements of the device were unable to be taken due to the condition of the device.The balloon was bunched over the tip.
|
| |
|
Event or Problem Description
|
|
It was reported that catheter entrapment and shaft break occurred.A 15mm x 5.00mm nc quantum apex balloon catheter was selected for use.However, upon entering a 0.14 unknown guide wire into the catheter, the balloon shaft was stuck on the wire in the guide catheter and as the physician was pushing the device, the catheter broke.The physician was able to remove everything as he pulled back the wire.A new wire and a 20x5.00mm balloon was used and completed the procedure.No patient complications were reported and the patient's status was fine.It was further reported that the target lesion was located in the mildly to moderately calcified vessel of the left leg.The device was fractured mid shaft and was removed with the guide wire and guide catheter.
|
| |
|
Additional Manufacturer Narrative
|
|
Describe event or problem updated.
|
| |
|
Event or Problem Description
|
|
It was reported that catheter entrapment and shaft break occurred.A 15mm x 5.00mm nc quantum apex balloon catheter was selected for use.However, upon entering a 0.14 unknown guide wire into the catheter, the balloon shaft was stuck on the wire in the guide catheter and as the physician was pushing the device, the catheter broke.The physician was able to remove everything as he pulled back the wire.A new wire and a 20x5.00mm balloon was used and completed the procedure.No patient complications were reported and the patient's status was fine.It was further reported that the target lesion was located in the mild to moderately calcified vessel of the left leg.The device was fractured mid shaft and was removed with the guide wire and guide catheter.
|
| |
|