DEPUY ORTHOPAEDICS INC US PINN SECTOR W/GRIPTION 62MM; PINNACLE HIP SYSTEM : HIP METAL ACETABULAR CUPS Back to Search Results
Model Number 1217-32-062
Medical Device Problem Codes Use of Device Problem (1670); Loss of Osseointegration (2408); Osseointegration Problem (3003)
Health Effect - Clinical Codes Inadequate Osseointegration (2646); No Code Available (3191)
Date of Event 12/03/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that the patient had acetabular cup revision due to loosening of the primary cup at the bone to implant interface.Revised to a 58 pinnacle multihole cup with three screws and a 36x58 neutral poly and +12 36mm femoral head.Doi: (b)(6) 2019; dor: (b)(6) 2019; left hip.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provision of 21 cfr, part 803.The report may be based on the information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: added: b5 and h6(device code).Corrected: h6(device code).Product complaint # :(b)(4).Investigation summary :no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.Device history lot :null.Device history batch :null.Device history review :null.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Additional information received indicating that examination of the provided x-ray image finds no evidence of device fracture, disassociation or anything indicative of a product problem.It is noted the acetabular cup position is both more vertical than would be recommended and also has less version than would be recommended.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PINN SECTOR W/GRIPTION 62MM
Common Device Name
PINNACLE HIP SYSTEM : HIP METAL ACETABULAR CUPS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key9503963
Report Number1818910-2019-124374
Device Sequence Number1839030
Product Code LPH
UDI-Device Identifier10603295010357
UDI-Public10603295010357
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
K071784
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2019
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/03/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1217-32-062
Device Catalogue Number121732062
Device Lot Number686181
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 12/03/2019
Supplement Date Received by Manufacturer01/20/2020
Initial Report FDA Received Date12/20/2019
Supplement Report FDA Received Date01/30/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ALTRX NEUT 36IDX62OD.; ARTICULEZE M HEAD 36MM +8.5.
Outcome Attributed to Adverse Event Required Intervention;
Patient Age75 YR
-
-