| Model Number |
1217-32-062 |
| Medical Device Problem Codes |
Use of Device Problem (1670); Loss of Osseointegration (2408); Osseointegration Problem (3003)
|
| Health Effect - Clinical Codes |
Inadequate Osseointegration (2646); No Code Available (3191)
|
| Date of Event |
12/03/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient had acetabular cup revision due to loosening of the primary cup at the bone to implant interface.Revised to a 58 pinnacle multihole cup with three screws and a 36x58 neutral poly and +12 36mm femoral head.Doi: (b)(6) 2019; dor: (b)(6) 2019; left hip.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provision of 21 cfr, part 803.The report may be based on the information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: added: b5 and h6(device code).Corrected: h6(device code).Product complaint # :(b)(4).Investigation summary :no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.Device history lot :null.Device history batch :null.Device history review :null.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Additional information received indicating that examination of the provided x-ray image finds no evidence of device fracture, disassociation or anything indicative of a product problem.It is noted the acetabular cup position is both more vertical than would be recommended and also has less version than would be recommended.
|
| |
|