Model Number 176647 |
Device Problems
Break (1069); Detachment of Device or Device Component (2907)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 12/09/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
According to the reporter, fan retractor broke during laparoscopic prostatectomy.Small pieces of plastic from the retractor fell inside the patient and these were removed immediately once identified by the surgeon.A new fan retractor was used to complete the case.There was no patient injury.
|
|
Manufacturer Narrative
|
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.A visual inspection of the returned product noted that the clevis fixture was broken on both sides of the product.A functional evaluation found that the device articulated properly and without difficulty.The device fanned out properly and without difficulty.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Replication of the reported condition may occur because of the instrument being forced beyond its indicated use due to applying excessive stress over the clevis.The plastic clevis was broken from the pin hole; which is consistent with the use of excessive force during the procedure, thus causing the unit clevis to break and the distal end of the handle to partially split open.The root cause of the observed damage was due to the product not being used as indicated which caused or contributed to the reported condition.No further actions have been deemed necessary at this time.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|