MAUDE Adverse Event Report: SMITH & NEPHEW, INC. BOX CHISEL; PRSTHSS,KNEE,PTLL/FMRTBL,SM-CNSTRND,UNCMNTD,PRS,CTD,PLYMR/MTL/PLYMR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. BOX CHISEL; PRSTHSS,KNEE,PTLL/FMRTBL,SM-CNSTRND,UNCMNTD,PRS,CTD,PLYMR/MTL/PLYMR Back to Search Results
Model Number 71440373
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 11/20/2019
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during tka box chisel is bent and it would not fit through the housing collet.Delay greater than 30 minutes reported, back-up device available and no injuries have been confirmed.
 
Additional Manufacturer Narrative
The device, used in treatment, was not returned for evaluation, the reported event could not be confirmed.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Without the actual product involved, our investigation cannot proceed.If the device or new information is received in the future, this complaint can be re-opened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BOX CHISEL
Common Device Name
PRSTHSS,KNEE,PTLL/FMRTBL,SM-CNSTRND,UNCMNTD,PRS,CTD,PLYMR/MTL/PLYMR
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key9504283
Report Number1020279-2019-04532
Device Sequence Number11232356
Product Code MBH
UDI-Device Identifier00885556034965
UDI-Public00885556034965
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K121393
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/10/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71440373
Device Catalogue Number71440373
Initial Date Received by Manufacturer 11/21/2019
Supplement Date Received by Manufacturer03/05/2020
Initial Report FDA Received Date12/20/2019
Supplement Report FDA Received Date03/10/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-