Model Number APTIO AUTOMATION
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Death (1802)
Date of Event 12/02/2019
Type of Reportable Event Death
Event or Problem Description
The customer alleged that there was a delay in obtaining a troponin (tnih) test result for a patient sample for approximately an hour on (b)(6) 2019.The customer reported that the patient expired while awaiting the troponin (tnih) result.The emergency room nurse followed up on the troponin results multiple times, making the operator aware of the delay.Statements and actions attributed to the customer are derived from information submitted to the siemens complaint handling system and haven't been verified.Siemens investigated the instrument files and determined that an initial panel of blood tests were ordered on the patient sample.A troponin test was subsequently ordered on the patient sample.Siemens determined that a sample tube with sid (b)(6) was returned to the aptio automation system (serial number: (b)(4)) on (b)(6) 2019 at 4:57:52 pm after being processed on an atellica instrument for the initial panel of results.The tube was sealed and sent to storage.At 6:41:48 pm, the sample was retrieved from storage, due to an add-on troponin (tnih) test that was subsequently ordered on the sample and processed at 6:48:55 pm.Then, the tube was sealed and sent to storage.At 7:38:43 pm, the sample was retrieved from storage and at 7:38:52 pm a de-sealer error sc021 "sample needs manual unseal - tube sealed too many times" was generated by the aptio automation system, as intended.The error caused the sample tube to route to a priority output lane for operator intervention at 7:39:30 pm, but was not reintroduced onto the track until 9:25:32 pm.The device is performing according to specifications, and there was no need for service or maintenance.Siemens is the importer of the medical device described.
 
Event or Problem Description
The initial mdr 3007494875-2019-00002 was filed on 20-dec-2019.Additional information (30-dec-2019): the manufacturer of the device submitted mdr 3010825766-2019-00014 on 30-dec-2019.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
APTIO AUTOMATION
Common Device Name
APTIO AUTOMATION
Manufacturer (Section D)
INPECO S.P.A
via torraccia 26
registration #: 3010825766
novazzano, 6883
SZ  6883
MDR Report Key9505082
Report Number3007494875-2019-00002
Device Sequence Number16604000
Product Code LXG
UDI-Device Identifier07640172342008
UDI-Public(01)07640172342008(11)190128(21)
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Importer
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/16/2020
Report Date (Section F) 12/20/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberAPTIO AUTOMATION
Device Catalogue Number11266636
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Distributor Facility Aware Date12/03/2019
Device Age4 YR
Event Location Hospital
Date Report to Manufacturer12/20/2019
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date12/20/2019
Supplement Report FDA Received Date01/16/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Death;
Patient Age81 YR
-
-