| Model Number |
LF5544 |
| Medical Device Problem Codes |
Peeled/Delaminated (1454); Device Damaged Prior to Use (2284); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
12/10/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, prior to use, the device's jaws started falling apart and the coating was burnt off.There was no patient involved.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: one device was received for evaluation.The returned product did not meet specification as received.Visual inspection found the clear boot was torn, all pieces were still attached and present.The jaws are also bent.The reported condition was confirmed.The device failed hipot testing.The welds on the jaws were broken.The jaws appeared to have been flexed too far.The investigation identified the root cause of the reported event to be user error.The instructions for use (ifu) states to prevent damage to the flexible insulation proximal to the jaws, confirm the handle is fully closed prior to insertion into and extraction from the cannula.Use the appropriately sized cannula to allow for easy insertion and extraction of the instrument.Failure to do so may impact the integrity of the flexible insulation.Cannulas with hard, non-beveled openings may cause the flexible insulation to retract, which may compromise the insulation.If retraction occurs, the instrument must be discarded.Do not attempt to clean the flexible insulation.Cleaning may damage insulation.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|