Lot Number 182101
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Code Available (3191)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Based on the review conducted on this case, lenstec was unable to perform a more thorough investigation of the complaint as the lens remains implanted.Therefore, we are unable to comment on the nature of the reported incident described and cannot corroborate the claim of exudative proliferative membrane or make any definitive statement as to its cause.Additionally, lenstec can confirm that all procedures in the manufacturing and packaging of the device were conducted correctly.Furthermore, the investigation and assessment from the medical monitor indicate there is no evidence to suggest that the incident was lens related.
 
Event or Problem Description
Lenstec received an email stating that "three patients had the symptoms of exudative proliferative membrane in front of the intraocular lens after implanting softec hdo.It was also noted that the doctor gave them hormone treatment'.All lenses remain implanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SOFTEC HDO
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
LENSTEC BARBADOS INC.
airport commercial centre
pilgrim road
christ church,
BB 
MDR Report Key9506925
Report Number9613160-2019-00036
Device Sequence Number11066521
Product Code HQL
UDI-Device Identifier00844369037365
UDI-Public00844369037365
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
P090022
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Reporter Type Manufacturer
Report Source distributor
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 12/20/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/05/2023
Device Lot Number182101
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/22/2019
Initial Report FDA Received Date12/20/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-