| Model Number |
400CE |
| Medical Device Problem Code |
Insufficient Heating (1287)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
12/06/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Gentherm received a report ((b)(4)) that the hemotherm heater/cooler would not heat above 34c.Unit was switched out and procedure was completed without further issue.No patient harm reported.Unit has not been returned for evaluation.
|
| |
|
Event or Problem Description
|
|
Customer reported that during the rewarming portion of a procedure, the hemotherm device would not heat above 34c.The unit was replaced and the procedure concluded without incident.No patient harm was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Device was evaluated and determined to have a defective heater.The heater was replaced to correct.
|
| |
|