Model Number MS1-4590S
Medical Device Problem Code Defective Device (2588)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 11/25/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
No product has been returned for investigation nor were x-rays or ultrasound images provided to confirm the alleged event.It is unknown if patient complied with post-operative physical restrictions.No root cause can be confirmed at this time.
 
Event or Problem Description
Information was received that a revision procedure was performed on (b)(6) 2019.As per the reporter the patient had a broken rod.
 
Additional Manufacturer Narrative
Investigation findings: product was returned and visual inspection was performed, customers alleged event was confirmed.Based on the investigation findings bending forces applied to the rod from patient¿s anatomy/activity may have caused the distraction rod to wedge into the housing tube and caused wear and tear, which would cause the rod to be jammed.Review of the device history record revealed no discrepancies related to this complaint.The rod was manufactured in accordance with the specified requirements and met all of the required quality inspections.
 
Event or Problem Description
This report has been updated to include investigation findings.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAGEC SPINAL BRACING AND DISTRACTION SYSTEM
Common Device Name
GROWING ROD SYSTEM-MAGNETIC ACTUATION
Manufacturer (Section D)
NUVASIVE SPECIALIZED ORTHOPEDICS, INC.
101 enterprise
suite 100
aliso viejo CA 92656
MDR Report Key9506953
Report Number3006179046-2019-00186
Device Sequence Number1878292
Product Code PGN
UDI-Device Identifier00812258026295
UDI-Public812258026295
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K171791
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2019
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/18/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMS1-4590S
Device Catalogue NumberPA0516
Device Lot Number8161028-02
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 12/06/2019
Supplement Date Received by Manufacturer03/25/2020
Initial Report FDA Received Date12/20/2019
Supplement Report FDA Received Date06/19/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age9 YR
Patient Weight32
-
-