| Model Number |
MS1-4590S |
| Medical Device Problem Code |
Defective Device (2588)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
11/25/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
No product has been returned for investigation nor were x-rays or ultrasound images provided to confirm the alleged event.It is unknown if patient complied with post-operative physical restrictions.No root cause can be confirmed at this time.
|
| |
|
Event or Problem Description
|
|
Information was received that a revision procedure was performed on (b)(6) 2019.As per the reporter the patient had a broken rod.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation findings: product was returned and visual inspection was performed, customers alleged event was confirmed.Based on the investigation findings bending forces applied to the rod from patient¿s anatomy/activity may have caused the distraction rod to wedge into the housing tube and caused wear and tear, which would cause the rod to be jammed.Review of the device history record revealed no discrepancies related to this complaint.The rod was manufactured in accordance with the specified requirements and met all of the required quality inspections.
|
| |
|
Event or Problem Description
|
|
This report has been updated to include investigation findings.
|
| |
|