Model Number 71335756
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Code Pain (1994)
Date of Event 04/11/2017
Type of Reportable Event Serious Injury
Event or Problem Description
*us legal*it was reported that the patient presented pain problems associated with a total hip replacement.A revision surgery was performed to explant the head and the liner.
 
Additional Manufacturer Narrative
The devices, intended for use in treatment, were not returned for evaluation and the reported event could not be confirmed.A clinical evaluation was conducted and the root cause of the reported clinical symptoms prior to the first revision cannot be confirmed, and it cannot be concluded that the reported reactions/events were associated with a mal-performance of the implant.However, the root cause of the second revision is due to the positioning of the acetabular component, and the long screw that was irritating the iliopsoas.The patient impact beyond the revision and associated pain cannot be determined.No further medical assessment is warranted at this time.A review of complaint history on the listed parts revealed no prior complaints for the listed batches with the same failure mode.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the risk management file and instructions for use documents revealed this failure mode was previously identified.Some potential probable causes of the reported pain could be device misalignment or migration.Based on this investigation, the need for corrective action is not indicated.Without the actual product involved and/or device information, our investigation cannot proceed.If the device or new information is received in the future, this complaint can be re-opened.No further actions are being taken at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
R3 20 DEG XLPE ACET LNR 36MM X 56MM
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER, POROUS
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key9507386
Report Number1020279-2019-04535
Device Sequence Number14198442
Product Code MBL
UDI-Device Identifier03596010598356
UDI-Public03596010598356
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K113848
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Device Explanted Year2017
Reporter Type Manufacturer
Report Source company representative,other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/03/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number71335756
Device Catalogue Number71335756
Device Lot Number16MM17078
Initial Date Received by Manufacturer 10/23/2019
Supplement Date Received by Manufacturer01/27/2020
Initial Report FDA Received Date12/20/2019
Supplement Report FDA Received Date02/03/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age53 YR
Patient Weight113
-
-