Model Number 406303
Medical Device Problem Code Component Misassembled (4004)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/04/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Recall number: 3005334138-12/5/2019-001-r.The investigation revealed that a 12f dilator and 12f sheath were packaged within the 14f fast-cath trio hemostasis introducer package.
 
Event or Problem Description
After a procedure the customer decided to return 129 unused devices for continued insertion difficulties.It was later determined that the introducers were 12f instead of the 14f as labeled.
 
Additional Manufacturer Narrative
16jan20 additional report required for received date.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FAST-CATH¿ TRIO HEMOSTASIS INTRODUCER, CATH-LOCK¿ LOCKING HUB, 12 CM SHEATH, 14
Common Device Name
INTRODUCER, CATHETER
Manufacturer (Section D)
ST. JUDE MEDICAL
5050 nathan lane north
plymouth MN 55442
MDR Report Key9507808
Report Number3005334138-2019-00971
Device Sequence Number2079880
Product Code DYB
UDI-Device Identifier05414734203906
UDI-Public05414734203906
Combination Product (Y/N)N
Initial Reporter CountryMY
PMA/510(K) Number
K061015
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 01/21/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2022
Device Model Number406303
Device Catalogue Number406303
Device Lot Number7133555
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 11/22/2019
Supplement Date Received by Manufacturer01/03/2020
01/03/2020
Initial Report FDA Received Date12/20/2019
Supplement Report FDA Received Date01/20/2020
01/21/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
SEE H10
Patient Sequence Number1
-
-