• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL - NEUROMODULATION EON MINI IPG, 16-CHANNEL RECHARGEABLE; SCS IPG

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL - NEUROMODULATION EON MINI IPG, 16-CHANNEL RECHARGEABLE; SCS IPG Back to Search Results
Model Number 3788
Device Problems Improper or Incorrect Procedure or Method (2017); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Irritation (1941); Inadequate Pain Relief (2388)
Event Date 01/01/2014
Event Type  Injury  
Manufacturer Narrative
Date of event¿ is estimated.The results/method and conclusion codes along with investigation results will be provided in the final report.
 
Event Description
It was reported that the patient felt tonic therapy to be irritating and wanted access to burst therapy.As a result, the patient stopped recharging the ipg as recommended.In turn, the ipg became inoperable.As a result, surgery occurred to explant and replace the ipg and the issue was resolved.
 
Manufacturer Narrative
The directions for use states that the implant must be charged every 30 to 90 days to avoid battery depletion.  based on the information received, the cause of the reported incident is related to user error.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EON MINI IPG, 16-CHANNEL RECHARGEABLE
Type of Device
SCS IPG
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer Contact
andrea deitz
6901 preston road
plano, TX 75024
9723098000
MDR Report Key9508564
MDR Text Key173983175
Report Number1627487-2019-13593
Device Sequence Number1
Product Code LGW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/01/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date11/28/2013
Device Model Number3788
Device Catalogue Number3788
Device Lot Number3546775
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/02/2019
Initial Date FDA Received12/21/2019
Supplement Dates Manufacturer Received01/30/2020
Supplement Dates FDA Received02/01/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/15/2011
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Weight100
-
-