|
Report confirmed.Evaluation determined that the leg was bent.It was also noted that the device failed the heat test and the temperature was measured to be 120°f.The bearings, spacer ring, o-ring, tactile feel ring and c-clip were worn and the color band was faded.Previous investigation performed under pr#127716 determined that the likely causes are debris in the collet and improper insertion of the tool.The user manual contains the following warnings ¿do not use a legend attachment if any part of the attachment appears to be bent, loose, missing or damaged.Do not use excessive pressure, such as bending or prying, on attachments or dissecting tools.This may cause tool to bend or break and cause injury to patient, operator and/or operating room staff." in addition, ¿insert dissecting tool into motor collet with a slight rotational motion.A tactile and audible click is observed indicating that the dissecting tool is fully seated.¿ the preventive maintenance/service manual for the legend system specifies service intervals for devices based on the hospital usage level.The maximum specified service interval is 24 months.Device has been in use for approximately 43 months with no record of factory service during this period.We will continue to track and trend this complaint type.Initial reporter: phone number: (b)(6).If information is provided in the future, a supplemental report will be issued.
|