OLYMPUS MEDICAL SYSTEMS CORP. EVIS EXERA II SMALL INTESTINAL VIDEOSCOPE; SMALL INTESTINAL VIDEOSCOPE Back to Search Results
Model Number SIF-Q180
Device Problem Microbial Contamination of Device (2303)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
The subject device in this report has not been returned to omsc for evaluation.The exact cause of the reported event could not be conclusively determined at this time.If additional information becomes available, this report will be supplemented.
 
Event Description
Olympus medical systems corp.(omsc) was informed that as a result of microbiological testing by the user facility, the subject device tested positive for unspecified microbes.The user facility did not provide other detailed information such as the number and the type of microbes.Other detailed information such as the reprocessing method was not provided.There was no report of infection associated with this report.
 
Manufacturer Narrative
This supplemental report is being submitted to provide additional information.The subject device has not been returned to omsc but was returned to olympus deutschland gmbh (ode).Ode sent the device to a third party laboratory for microbiological testing.As a result of the testing, no microbe was detected from the sample collected from the distal end unit, the instrument channel, and the air/water channel of the subject device.The testing result cleared the german guideline.The exact cause of the reported event could not be conclusively determined at this time.If additional information becomes available, this report will be supplemented.
 
Manufacturer Narrative
This supplemental report is being submitted to provide additional information.Omsc reviewed the manufacturing history (dhr) of the subject device and confirmed no irregularity.The exact cause of the reported event could not be conclusively determined.If additional information becomes available, this report will be supplemented.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA II SMALL INTESTINAL VIDEOSCOPE
Type of Device
SMALL INTESTINAL VIDEOSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key9508699
MDR Text Key205111001
Report Number8010047-2019-04519
Device Sequence Number1
Product Code FDA
UDI-Device Identifier04953170339936
UDI-Public04953170339936
Combination Product (y/n)N
PMA/PMN Number
K071254
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Type of Report Initial,Followup,Followup
Report Date 05/24/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberSIF-Q180
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/04/2019
Initial Date FDA Received12/22/2019
Supplement Dates Manufacturer Received12/20/2019
04/28/2020
Supplement Dates FDA Received01/07/2020
05/25/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-