A complete analysis and testing of the insulin pump showed that it was functioning properly and passed all functional testing.After testing it was concluded that the device operated within specifications.All buttons functioning properly.No damage/unlocked j2/lcd keypad connector during visual inspection noted.Device was received with normal operating currents and passed self-test, unexpected restart error test.No unexpected low battery, batt out limit or failed batt test alarms during testing.Also, device passed rewind test, basic occlusion test, occlusion test, prime/compromised force sensor system test, excessive no delivery test and displacement test.No unexpected no delivery alarm or rewind louder anomaly noted during testing.Device had minor scratched lcd window, cracked lcd window, cracked case at display window corners.The asr exemption e2015043 was revoked on february 4, 2019.This event was previously reported as an asr.Additional information or analysis results regarding this event will be reported as an mdr.
|