• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US QUICKSET ACE GRATER HEAD 51MM; HIP INSTRUMENTS : REAMERS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US QUICKSET ACE GRATER HEAD 51MM; HIP INSTRUMENTS : REAMERS Back to Search Results
Catalog Number 244000551
Device Problem Break (1069)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/02/2019
Event Type  malfunction  
Manufacturer Narrative
Product complaint #: (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
The piece broke before use on patient in preparation for thr surgery.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provision of 21 cfr, part 803.The report may be based on the information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint # (b)(4).Investigation summary: the product was returned for examination and the complainants findings confirmed.The grater shell section had detached from the base.The edge of the shell showed signs of excessive wear.The root cause was deemed to be wear and tear.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot: null.Device history batch: null.Device history review: null.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUICKSET ACE GRATER HEAD 51MM
Type of Device
HIP INSTRUMENTS : REAMERS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key9510936
MDR Text Key184583054
Report Number1818910-2019-124536
Device Sequence Number1
Product Code HTO
UDI-Device Identifier10603295123989
UDI-Public10603295123989
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 12/03/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number244000551
Device Lot NumberA1205
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/17/2019
Initial Date Manufacturer Received 12/03/2019
Initial Date FDA Received12/23/2019
Supplement Dates Manufacturer Received01/24/2020
Supplement Dates FDA Received01/27/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-