• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P. EPIK KNEE; BASEPLATE, ONLAY TIBIA, SIZE45 RM/LL EPIK

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. EPIK KNEE; BASEPLATE, ONLAY TIBIA, SIZE45 RM/LL EPIK Back to Search Results
Model Number 300-02-045
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 11/26/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Additional reporting on this event will be provided as a supplemental report to this document as soon as it becomes available.
 
Event or Problem Description
Revision surgery - conversion of total knee.
 
Additional Manufacturer Narrative
Manufacturer narrative: the reason for this revision surgery was due to patient needing to be converted to a total knee.The previous surgery and the surgery detailed in this event occurred 2 years and 1 months apart.There was no information submitted with this complaint about any patient activities, accidents, or medical contraindications that may have contributed to the reported event.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The surgery was completed as intended and without incident.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the device history record (dhr) show that the reported components used in the previous surgery, when released for use, met design and manufacturing requirements.Firstly there was a non conformance in main part#300-02-042, baseplate, onlay tibia, size45 rm/ll epik, which documents out of 8 parts lot 2 parts was rejected due to out of tolerance.All other items in the lot were met with fit, form and function.Secondly, there was another non conformance in concomitant part#340-45-009, insert, onlay tibial, epik, size45x9, which documents out of 10 parts lot 1 part was rejected due to out of tolerance.All other items in the lot were met with fit, form and function.The devices were verified to have gone through an acceptable sterilization process and were within its expiration date at the time of the previous surgery.Customer complaint history of the reported devices showed no present trends or on-going issues that are needing a review.The root cause of this complaint was a revision surgery due to positioning of cup and liner was off.There were no findings during this evaluation that indicate the reported devices were defective.There are multiple factors that may contribute to an event that are outside the control of djo surgical such as poor bone density, patient bone deterioration, inadequate soft tissue support, excessive range of motion, patient activities or trauma.There were no indication of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EPIK KNEE
Common Device Name
BASEPLATE, ONLAY TIBIA, SIZE45 RM/LL EPIK
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin, tx 78758-5445
MDR Report Key9511004
Report Number1644408-2019-01268
Device Sequence Number2777434
Product Code HRY
UDI-Device Identifier00888912113380
UDI-Public(01)00888912113380
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K022437
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2019
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 02/01/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/31/2020
Device Model Number300-02-045
Device Catalogue Number300-02-045
Device Lot Number082B1006
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 11/26/2019
Supplement Date Received by Manufacturer01/02/2020
Initial Report FDA Received Date12/23/2019
Supplement Report FDA Received Date02/01/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
340-45-009 LOT 334B1038
Outcome Attributed to Adverse Event Required Intervention;
Patient Age59 YR
-
-