|
Allergan did not submit this mdr within 30 days of becoming aware.Recent stimulated reporting related to 2011068-7/2/19-001-r has increased complaint and mdr volume.Allergan is implementing a plan to address the increased volumes.A review of the device history record has been initiated.If any new, changed or corrected information is noted, a supplemental medwatch will be submitted.Device evaluation: visual analysis of the returned device identified: weight to the spec, crease fold, deformation.Cloudy and bubbles were observed after autoclave disinfection process.Based on the device analysis the final assessment is: no issues found related with the manufacturing process.The event of "capsular contracture" is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Reason for reoperation: capsular contracture.
|