BIOMET 3I 3I T3® NON-PLATFORM SWITCHED PARALLEL WALLED IMPLANT 3.25 X 8.5MM; DENTAL IMPLANT Back to Search Results
Model Number BOSS385
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 11/26/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Zimmer biomet complaint (b)(4).Patient's weight not provided/unknown.A summary investigation has been completed for lack of primary stability events that do not allege a deficiency with the implant and identified that the reported event is likely due to biological factors which have an adverse effect on implant stability or is related to surgical technique and insertion torque.Due to a wide range of external factors (non-design/ non-manufacturing related), identifying a definitive root cause is generally not possible.Should additional information be received which indicates that the device may have caused or contributed to the event, an additional report will be submitted.Summary investigation.
 
Event or Problem Description
It was reported that the implant could not achieve primary stability.The implant was removed and no other implant was placed.Bone loss was also reported.
 
Additional Manufacturer Narrative
The event was updated stating that it was poor bone quality that caused the implant to not be able to achieve primary stability.Bone loss was removed from the event description.
 
Event or Problem Description
It was reported that the implant could not achieve primary stability.The implant was removed and no other implant was placed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3I T3® NON-PLATFORM SWITCHED PARALLEL WALLED IMPLANT 3.25 X 8.5MM
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
MDR Report Key9513029
Report Number0001038806-2019-01688
Device Sequence Number366349
Product Code DZE
UDI-Device Identifier00844868028802
UDI-Public(01)00844868028802(17)230302(1
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K122300
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date (Section B) 02/06/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/02/2023
Device Model NumberBOSS385
Device Catalogue NumberBOSS385
Device Lot Number2018021129
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/29/2019
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 11/27/2019
Supplement Date Received by Manufacturer01/07/2020
Initial Report FDA Received Date12/23/2019
Supplement Report FDA Received Date02/06/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;