| Model Number |
S1PIC1.9-C |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Cardiac Tamponade (2226)
|
| Date of Event |
11/08/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The suspect device was returned for investigation.Catheter was received without any apparent signs of damage.Inspection of the returned catheter revealed that there was no device failure and that the device functioned as intended.The device history record for lot #180828 was reviewed.All catheters passed in-process and final inspection criteria including leak testing.Cardiac tamponade is a known inherit risk during the use of peripherally inserted central catheters and most commonly attributed to a malpositioned catheter tip.
|
| |
|
Event or Problem Description
|
|
Infant developed cardiac tamponade after about 6 hours of line being placed.
|
| |
|