• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENDOLOGIX SANTA ROSA OVATION IX; MAIN BODY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENDOLOGIX SANTA ROSA OVATION IX; MAIN BODY Back to Search Results
Model Number TV-AB2980-J
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 11/25/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device involved in this event will not be returned for evaluation as it remains implanted in the patient.Patient medical records and imaging studies will be requested for further evaluation by the clinical specialist.If additional information pertinent to the incident is obtained, a follow-up report will be submitted.Device remains implanted.
 
Event or Problem Description
The patient was initially implanted with the ovation ix abdominal stent graft system to treat an abdominal aortic aneurysm (aaa).Approximately two and half (2.5) years post initial procedure, the patient presented with an endoleak of indeterminate origin (suspected type 2).The patient is currently being monitored.Note: this patient had a previous report of a type 2 endoleak (non device related failure) in (b)(6) of 2018.
 
Additional Manufacturer Narrative
Clinical assessment of the reported event was unable to be completed due to a lack of receipt of relevant medical records and/or imaging.Only a discharge summary was received with little relevant information to this case.Requests for medical images were made; however, no response was received from the hospital.As such, event determination, off label conditions, related patient harms and patient disposition could not be assessed.The manufacturing lot evaluation confirmed all devices met specifications prior to release.The device remains implanted; therefore, no device evaluation was completed.No additional investigation of this reported event is planned; however, if additional information pertinent to the incident is obtained, a follow-up report will be submitted.Endologix will continue to monitor this complaint and similar complaints in the event further investigation is needed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OVATION IX
Common Device Name
MAIN BODY
Manufacturer (Section D)
ENDOLOGIX SANTA ROSA
3910 brickway blvd
santa rosa CA 95403
MDR Report Key9513449
Report Number3008011247-2019-00183
Device Sequence Number1410089
Product Code MIH
UDI-Device IdentifierM701TVAB2980J1
UDI-Public+M701TVAB2980J1/$$3200118FS120916302
Combination Product (Y/N)N
Initial Reporter StatePA
PMA/510(K) Number
P120006
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Remedial Action Other
Type of Report Initial,Followup
Report Date (Section B) 11/25/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/18/2020
Device Model NumberTV-AB2980-J
Device Catalogue NumberTV-AB2980-J
Device Lot NumberFS120916-30
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 11/25/2019
Supplement Date Received by Manufacturer11/25/2019
Initial Report FDA Received Date12/23/2019
Supplement Report FDA Received Date07/29/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
FILL POLYMER, LOT # FF041817-01.; ILIAC LIMB, LOT # FS011317-05.; ILIAC LIMB, LOT # FS120616-47.
Outcome Attributed to Adverse Event Other;
Patient Age86 YR
-
-