Patient information was unavailable from the site.A medtronic representative went to the site to test the equipment.The internal monitor cable, external monitor cable and surgeon monitor were replaced.The system then passed the system checkout and was found to be fully functional.The internal monitor cable was returned to the manufacturer for analysis.The internal monitor cable was found to be fully functional with no problem found.The reported event could not be duplicated by medtronic personnel.The external monitor cable was returned to the manufacturer for analysis.The external monitor cable was found to be fully functional with no problem found.The reported event could not be duplicated by medtronic personnel.The surgeon monitor was returned to the manufacturer for analysis.The surgeon monitor was found to be fully functional with no problem found.The reported event could not be duplicated by medtronic personnel.If information is provided in the future, a supplemental report will be issued.
|