• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. R3 0 DEG XLPE ACET LNR 36MM X MM56; PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL POROUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. R3 0 DEG XLPE ACET LNR 36MM X MM56; PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL POROUS Back to Search Results
Model Number 71332756
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Injury (2348)
Date of Event 11/25/2019
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient has a polarstem device in place since 2017 and was experiencing elevated cobalt levels.Revision surgery was performed and the poly was explanted and an oxiniu head was used.
 
Additional Manufacturer Narrative
The devices, used in treatment, were not returned for evaluation and the reported event could not be confirmed.The medical investigation concluded that, per complaint details, a polarstem was revised (b)(4) approximately 2 years post implantation along with a cocr femoral head (b)(4) and xlpe liner exchange (b)(4) due to elevated cobalt levels.No clinically relevant medical documentation was provided; therefore, a thorough medical investigation could not be performed.The root cause could not be confirmed and the patient impact beyond the reported elevated metal ions and revision procedure could not be determined.No further medical assessment is warranted at this time.A review of complaint history did not reveal additional complaints for the listed batches.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the risk management file and instructions for use documents revealed this failure mode was previously identified.Some potential probable causes for this event could include lifetime of the device or joint tightness.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the devices or additional information be received, the complaints will be reopened.No further investigation warranted for these complaints; however, we will continue to monitor for future complaints and investigate as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
R3 0 DEG XLPE ACET LNR 36MM X MM56
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL POROUS
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key9513573
Report Number1020279-2019-04570
Device Sequence Number2777475
Product Code MBL
UDI-Device Identifier03596010598004
UDI-Public03596010598004
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K113848
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,consum
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/18/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device No Information
Device Expiration Date12/13/2024
Device Model Number71332756
Device Catalogue Number71332756
Device Lot Number14MM12431
Initial Date Received by Manufacturer 11/25/2019
Supplement Date Received by Manufacturer03/13/2020
Initial Report FDA Received Date12/23/2019
Supplement Report FDA Received Date03/18/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age77 YR
-
-