Primary gamma nail procedure.It was reported that outside of the patient, the targeter aligned to the dummy short gamma nail.In the patient, the targeter would not align for the distal screw on the short gamma nail.The case was converted to a long gamma nail so that the screws could be placed manually.Surgery was completed successfully with a delay of approximately 15 minutes.The wasted nail and targeter are allegedly available for return.
|
Primary gamma nail procedure.It was reported that outside of the patient, the targeter aligned to the dummy short gamma nail.In the patient, the targeter would not align for the distal screw on the short gamma nail.The case was converted to a long gamma nail so that the screws could be placed manually.Surgery was completed successfully with a delay of approximately 15 minutes.The wasted nail and targeter are allegedly available for return.
|
The reported event could not be confirmed, since the returned device is conforming to specifications and fully functional.The device inspection revealed the following: reveals no significant damage to devices.Functional inspection: devices were tested in complete assembly with the following: drill, drill sleeve, tissue protection sleeve, speedlock sleeve gamma 3 and gamma 3 nail ¿ 180mm.Devices were successfully assembled and drill bit seamlessly fit through all components as required and described in the optech without misalignment.Dimensional inspection: due to the nature of the complaint, a dimensional inspection was not considered necessary.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.Based on investigation, the root cause was attributed to a user related issue.The failure was caused by inadequate assembly of required devices.If any further information is provided, the complaint report will be updated.
|