LIVANOVA DEUTSCHLAND GMBH CENTRIFUGAL PUMP CONSOLE; PUMP, BLOOD, CARDIOPULMONARY BYPASS, NON-ROLLER TYPE Back to Search Results
Model Number 60-03-75
Device Problem Communication or Transmission Problem (2896)
Patient Problem No Patient Involvement (2645)
Event Date 11/25/2019
Event Type  malfunction  
Manufacturer Narrative
There was no patient involvement.Livanova (b)(4) manufactures the cenrifugal pump console.The incident occurred in (b)(6).A livanova field service representative was dispatched to the facility to investigate and could confirm the reported issue.He replaced the pressure module and the problem was solved.Subsequent functional verification testing was completed without further issues and the unit was returned to service.Livanova requested the part for further investigation but the customer retained the module and was not available to be returned.A defect of the pressure module led to the reported failure can be the most likely root cause.A review of the dhr could not identify any deviations or nonconformities relevant to the issue.
 
Event Description
Livanova received a report that the error message " no com!" associated to the pressure module was displayed on a centrifugal pump console during setup.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTRIFUGAL PUMP CONSOLE
Type of Device
PUMP, BLOOD, CARDIOPULMONARY BYPASS, NON-ROLLER TYPE
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND GMBH
lindberghstr. 25
munich 80309
GM  80309
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND GMBH
lindberghstr. 25
munich 80309
GM   80309
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key9514797
MDR Text Key196751363
Report Number9611109-2019-01000
Device Sequence Number1
Product Code KFM
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K020571
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial
Report Date 12/24/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/24/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number60-03-75
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received11/26/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/21/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse