Model Number 10-85-00 |
Device Problem
Pumping Stopped (1503)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 11/14/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
There was no patient involvement.Livanova (b)(4) manufactures the s5 double head pump.The incident occurred in (b)(6).A livanova field service representative was dispatched to the facility to investigate and replaced the motor control board.The issue was solved.Subsequent functional verification testing was completed without further issues and the unit was returned to service.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
|
Event Description
|
Livanova (b)(4) received a report that a s5 double head pump stopped and showed an error message during service.There was no report of patient injury.
|
|
Manufacturer Narrative
|
H.10: the replaced board was requested back to the manufacturer site for a detailed investigation.No defects and no functional issues could be found on the board.It is reasonable to assume that the reported issue was caused by a loose connection.
|
|
Event Description
|
See initial report.
|
|