MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION MAXFORCE; CATHETER, BILIARY, DIAGNOSTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION MAXFORCE; CATHETER, BILIARY, DIAGNOSTIC Back to Search Results
Model Number M00567370
Medical Device Problem Code Material Puncture/Hole (1504)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 11/26/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Although the suspect device has been received, the evaluation has not been completed.Therefore, the cause of the reported malfunction has not been determined.Upon completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental mdr will be filed.
 
Event or Problem Description
It was reported to boston scientific corporation that a maxforce biliary dilatation balloon was used in the hepatic portal during a biliary stent placement procedure performed on (b)(6) 2019.According to the complainant, during the procedure, it was noted that the balloon had a pinhole-like hole.The procedure was completed with another maxforce biliary dilatation balloon.There have been no patient complications reported as a result of this event.
 
Additional Manufacturer Narrative
Block h6: problem code 1504 captures the reportable event of balloon pinhole.Block h10: investigation results a visual examination of the returned complaint device found that the balloon did not have any visual defects and lookled normal.The inner radio opaque (ro) markers inside the balloon were inspected for any sharp edge but none of them showed abnormalities.Functional analysis was performed, and the balloon was inflated without a problem; however, the balloon would not hold pressure due to a pinhole in the body of the balloon.This failure is likely due to factors or conditions related to the procedure that could have affected its performance and its intended purpose, such as the technique used by the physician during the procedure, the amount of strength applied by the customer during the movement of the balloon and/or the interaction between the scope and the balloon.Therefore, the most probable root cause is adverse event related to procedure.A review of the device history record (dhr) was performed and confirmed that this device met all material, assembly and performance specifications.
 
Event or Problem Description
It was reported to boston scientific corporation that a maxforce biliary dilatation balloon was used in the hepatic portal during a biliary stent placement procedure performed on (b)(6) 2019.According to the complainant, during the procedure, it was noted that the balloon had a pinhole-like hole.The procedure was completed with another maxforce biliary dilatation balloon.There have been no patient complications reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAXFORCE
Common Device Name
CATHETER, BILIARY, DIAGNOSTIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
MDR Report Key9515931
Report Number3005099803-2019-06185
Device Sequence Number11233632
Product Code FGE
UDI-Device Identifier08714729097068
UDI-Public08714729097068
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K910931
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/11/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/20/2022
Device Model NumberM00567370
Device Catalogue Number6737
Device Lot Number0023826070
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/18/2019
Initial Date Received by Manufacturer 11/27/2019
Supplement Date Received by Manufacturer01/16/2020
Initial Report FDA Received Date12/24/2019
Supplement Report FDA Received Date02/11/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-