| Model Number |
M00567370 |
| Medical Device Problem Code |
Material Puncture/Hole (1504)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
11/26/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Although the suspect device has been received, the evaluation has not been completed.Therefore, the cause of the reported malfunction has not been determined.Upon completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental mdr will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that a maxforce biliary dilatation balloon was used in the hepatic portal during a biliary stent placement procedure performed on (b)(6) 2019.According to the complainant, during the procedure, it was noted that the balloon had a pinhole-like hole.The procedure was completed with another maxforce biliary dilatation balloon.There have been no patient complications reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Block h6: problem code 1504 captures the reportable event of balloon pinhole.Block h10: investigation results a visual examination of the returned complaint device found that the balloon did not have any visual defects and lookled normal.The inner radio opaque (ro) markers inside the balloon were inspected for any sharp edge but none of them showed abnormalities.Functional analysis was performed, and the balloon was inflated without a problem; however, the balloon would not hold pressure due to a pinhole in the body of the balloon.This failure is likely due to factors or conditions related to the procedure that could have affected its performance and its intended purpose, such as the technique used by the physician during the procedure, the amount of strength applied by the customer during the movement of the balloon and/or the interaction between the scope and the balloon.Therefore, the most probable root cause is adverse event related to procedure.A review of the device history record (dhr) was performed and confirmed that this device met all material, assembly and performance specifications.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that a maxforce biliary dilatation balloon was used in the hepatic portal during a biliary stent placement procedure performed on (b)(6) 2019.According to the complainant, during the procedure, it was noted that the balloon had a pinhole-like hole.The procedure was completed with another maxforce biliary dilatation balloon.There have been no patient complications reported as a result of this event.
|
| |
|