COOK ENDOSCOPY TRI-TOME PC PROTECTOR; KNS Back to Search Results
Catalog Number TRI-25M-P
Device Problem Failure to Align (2522)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/28/2019
Event Type  malfunction  
Manufacturer Narrative
Investigation evaluation: our evaluation of the returned device confirmed the report of incorrect cutting wire orientation.During the laboratory analysis, the sphincterotome was advanced through a duodenoscope that is placed in a simulated biliary position.The duodenoscope has an accessory channel that is 4.2 mm in diameter.The catheter exited the endoscope with the cutting wire facing 2 o'clock.The distal end was advanced into the simulated papilla and bowed and the cutting wire oriented in the 1 o'clock position.The sphincterotome catheter was subjected to a close visual examination and a small amount of twisting was observed at the distal end.A discrepancy or anomaly that could have contributed to the reported event was not found during our laboratory analysis of the returned product.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: a definitive cause for this observation could not be determined because the actual use conditions could not be duplicated in the laboratory setting.Due to a variety of clinical conditions such as patient anatomy, endoscope position or progression of disease state, we could not reproduce the actual conditions of product usage during our laboratory analysis.This limits our ability to conclusively determine a cause.Prior to distribution, all tri-tome pc protectors are subjected to a visual inspection and functional test to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: a review of the complaint history was conducted.The likelihood of occurrence is considered rare.Corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
 
Event Description
During an endoscopic procedure, the physician used a cook tri-tome pc protector.The device was rotated to three (3) o'clock [incorrect cutting wire orientation] and could not be used.A section of the device did not remain inside the patient's body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRI-TOME PC PROTECTOR
Type of Device
KNS
Manufacturer (Section D)
COOK ENDOSCOPY
4900 bethania station road
winston-salem NC 27105
Manufacturer (Section G)
COOK ENDOSCOPY
4900 bethania station road
winston-salem NC 27105
Manufacturer Contact
scottie fariole
4900 bethania station road
winston-salem, NC 27105
MDR Report Key9516221
MDR Text Key220061305
Report Number1037905-2019-00794
Device Sequence Number1
Product Code KNS
Combination Product (y/n)N
Reporter Country CodeKS
PMA/PMN Number
K901443
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 11/28/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/24/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/30/2022
Device Catalogue NumberTRI-25M-P
Device Lot NumberW4243520
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/12/2019
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/28/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/07/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-