• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HOUSTON CENTURION SURGICAL PROCEDURE PAK; GENERAL SURGERY TRAY (KIT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HOUSTON CENTURION SURGICAL PROCEDURE PAK; GENERAL SURGERY TRAY (KIT) Back to Search Results
Catalog Number 8065752200
Device Problem Break (1069)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/03/2019
Event Type  malfunction  
Manufacturer Narrative
Investigation, including root cause analysis, is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event Description
A customer reported that the balanced tip was broken during a cataract procedure.The product was replaced and the procedure was completed.There was no patient harm.
 
Manufacturer Narrative
Additional information provided in h.3, h.6 and h.10.A sample was not received at the manufacturing site for evaluation; however, the provided customer photo confirms the reported issue of broken phaco tip.A review of the device history records traceable to the reported lot number, indicates that the product was processed and released according to the product¿s acceptance criteria.The report of broken phaco tip is confirmed based on the photo; however, because a sample was not received at the manufacturing site and the device history record review of the lot number provided indicated the product was released to the product¿s acceptable criteria, the root cause for customer complaint issue cannot be determined.The exact root cause for this complaint is unknown; therefore, specific action with regards to this complaint cannot be taken.All phaco tips are 100% visually inspected by trained operators using 30x magnification during the manufacturing process.Complaints are reviewed and monitored at regular intervals for any significant adverse trends.No additional action is required at this time.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTURION SURGICAL PROCEDURE PAK
Type of Device
GENERAL SURGERY TRAY (KIT)
Manufacturer (Section D)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
MDR Report Key9516290
MDR Text Key177981030
Report Number1644019-2019-00205
Device Sequence Number1
Product Code LRO
Combination Product (y/n)N
PMA/PMN Number
K880961
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 03/19/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/24/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/30/2021
Device Catalogue Number8065752200
Device Lot Number2299943H
Was Device Available for Evaluation? No
Date Manufacturer Received03/05/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-