| Model Number |
R SERIES |
| Medical Device Problem Code |
Defibrillation/Stimulation Problem (1573)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The customer's report was duplicated and the pace/defib engine board was replaced to resolve the malfunction.The device was recertified and returned to the customer.Analysis of reports of this type has not identified an increase in trend.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during a routine shift check by a clinician, the device displayed a "defib disabled" message.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|
Search Alerts/Recalls
|