As the lot number for the device was provided, a manufacturing review will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.(expiry date 01/2022).Section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
|
It was reported that during a stent placement procedure in the mesenteric artery, the healthcare provider (hcp), aware of the off-label indication for use, predilated the lesion and advanced the balloon expandable vascular covered stent when it became stuck.It was further reported that upon attempt to pull it back, the stent allegedly dislodged from the balloon while inside the sheath.The delivery system and introducer sheath were removed.The procedure was completed by performing angioplasty.There was no reported patient injury.
|