| Model Number |
10603 |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
11/27/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device is a combination product.
|
| |
|
Event or Problem Description
|
|
It was reported that stent damage occurred.A 3.00 x 20mm synergy ii drug-eluting stent was selected for use.However, during preparation, it was noticed that the stent was flared.The device was never used and the procedure was completed without patient complications.
|
| |
|
Additional Manufacturer Narrative
|
|
Device is a combination product.Device evaluated by mfr.: synergy ii us mr 3.00 x 20mm stent delivery system was returned for analysis.A visual and microscopic examination of the stent found proximal stent damage, with stent struts lifted and pulled distally.The undamaged stent outer diameter was measured and the result is within maximum crimped stent profile measurement.The balloon cones were reviewed and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and microscopic examination of the bumper tip found distal tip damage.A visual and tactile examination of the hypotube found multiple hypotube kinks at several locations along the hypotube shaft.A visual examination of the outer and inner lumen and mid-shaft section found no issues along the shaft polymer extrusion.No other issues were identified during the product analysis.
|
| |
|
Event or Problem Description
|
|
It was reported that stent damage occurred.A 3.00 x 20mm synergy ii drug-eluting stent was selected for use.However, during preparation, it was noticed that the stent was flared.The device was never used and the procedure was completed without patient complications.
|
| |
|
Search Alerts/Recalls
|