• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC MEXICO ARTHROSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC MEXICO ARTHROSCOPE Back to Search Results
Catalog Number K09A
Medical Device Problem Codes Leak/Splash (1354); Material Rupture (1546)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 11/26/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Neither the device nor films of applicable imaging studies were returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that patient underwent unknown surgery.Intra-op, balloon ruptured during inflation.The contrast media leaked and the pressure (psi value) became zero(as seen in the monitor).Therefore, another product was used.Patient complication are unknown at this time.
 
Additional Manufacturer Narrative
Product analysis: visual and functional inspection revealed one of the balloons has been ruptured and the other one functions as intended.The rupture is a pinhole near the proximal peak.This rupture is consistent with the balloon coming in contact with bone spurs when the balloon is inflated in the vertebral body.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARTHROSCOPE
Common Device Name
ARTHROSCOPE
Manufacturer (Section D)
MEDTRONIC MEXICO
av. paseo del cucapah #10510
tijuana,bc 22570
MX  22570
MDR Report Key9517650
Report Number9612164-2019-05294
Device Sequence Number10885494
Product Code HRX
UDI-Device Identifier00763000025595
UDI-Public00763000025595
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K041454
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/19/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberK09A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/18/2020
Initial Date Received by Manufacturer 11/26/2019
Supplement Date Received by Manufacturer06/17/2020
Initial Report FDA Received Date12/25/2019
Supplement Report FDA Received Date06/19/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-