| Catalog Number |
KE152 |
| Medical Device Problem Codes |
Inflation Problem (1310); Device Slipped (1584)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
11/25/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Neither the device nor films of applicable imaging studies were returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient underwent unknown surgery at l1-l2 due to 2 level compression fracture.Intra-op, the balloon would not expand initially because the band on the proximal portion of the balloon that has the size of the kit on it (15/2) was loose.It slides up and down shaft of the balloon and could accidentally be implanted in the patient.It was not noticed and the balloon went up to 700 psi.Later it slipped out of the band and expanded rapidly.There was delay in overall procedure time and the procedure was successfully completed with the original product.There were no patient symptoms or complication reported as a result of this event.
|
| |
|