DAVIS & GECK CARIBE LTD VERSAPORT; LAPAROSCOPE, GENERAL & PLASTIC SURGERY Back to Search Results
Model Number 179093F
Device Problem Difficult to Remove (1528)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/11/2019
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, pre-operatively, on a laparoscopic hysteromyoma, during the preparation process, after using the suction catheter to enter and exit the two trocars for a while, the doctor felt that the suction catheter was stuck and unable to enter and exit.They opened a new product to complete the case.There was no patient injury.
 
Manufacturer Narrative
Additional information: d4(udi), d10, g4, h3, h6 h3 evaluation summary: post market vigilance (pmv) led an evaluation of two devices with fixation cannula.The visual inspection of the returned product noted: that the obturators were received inserted into the cannulas.Prolonged insertion cause the duckbill seals to become stretched.The obturators were removed and it was confirmed that the duckbill seals were stretched.The trocars, cannulas, obturators and circular seals appeared intact.A functional evaluation found that the devices failed an air leak test.The obturators were inserted into the cannula and removed cleanly.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.The investigation concluded there were no abnormalities that would have caused or contributed to the reported condition.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERSAPORT
Type of Device
LAPAROSCOPE, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DO  0101
MDR Report Key9519534
MDR Text Key181553219
Report Number9612501-2019-02518
Device Sequence Number1
Product Code GCJ
UDI-Device Identifier20884521081250
UDI-Public20884521081250
Combination Product (y/n)N
PMA/PMN Number
K062326
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 04/06/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/26/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/29/2024
Device Model Number179093F
Device Catalogue Number179093F
Device Lot NumberJ9C3059Y
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/16/2020
Date Manufacturer Received03/23/2020
Is This a Reprocessed and Reused Single-Use Device? No