• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: US SURGICAL PUERTO RICO SPACEMAKER PLUS; LAPAROSCOPE, GENERAL & PLASTIC SURGERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

US SURGICAL PUERTO RICO SPACEMAKER PLUS; LAPAROSCOPE, GENERAL & PLASTIC SURGERY Back to Search Results
Model Number SMSBTOVL
Medical Device Problem Code Inflation Problem (1310)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 12/05/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during left inguinal hernia surgery, the balloon did not inflate during surgery.Another balloon was used to complete the case.There was no patient injury.
 
Additional Manufacturer Narrative
Evaluation summary: post market vigilance (pmv) led an evaluation of two devices.The visual inspection of the returned products noted that the obturators, lock collars, trocar balloons and dissector balloons were received and appeared to be intact.The inflation syringes were received.The insufflation bulbs were received.The circular seal for the dissector balloons appeared intact.A functional evaluation found that the trocar balloons were inflated with no leaks detected.The dissector balloons were inflated with no leaks detected.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.The investigation concluded there were no abnormalities that would have caused or contributed to the reported condition.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPACEMAKER PLUS
Common Device Name
LAPAROSCOPE, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
US SURGICAL PUERTO RICO
201 sabanetas industrial park
ponce PR 00716 4401
MDR Report Key9519712
Report Number2647580-2019-06314
Device Sequence Number2851558
Product Code GCJ
UDI-Device Identifier10884521067820
UDI-Public10884521067820
Combination Product (Y/N)N
Initial Reporter StateNE
Initial Reporter CountryUS
PMA/510(K) Number
K042412
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source user facility
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 02/18/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2021
Device Model NumberSMSBTOVL
Device Catalogue NumberSMSBTOVL
Device Lot NumberP9G1231Y
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/10/2020
Initial Date Received by Manufacturer 12/12/2019
Supplement Date Received by Manufacturer02/04/2020
Initial Report FDA Received Date12/26/2019
Supplement Report FDA Received Date02/18/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age69 YR
Patient Weight72
-
-