| Model Number |
EEAXL2135 |
| Medical Device Problem Code |
Failure to Form Staple (2579)
|
| Health Effect - Clinical Codes |
Tissue Damage (2104); No Code Available (3191)
|
| Date of Event |
12/11/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during a laparoscopic gastric bypass procedure, after the staple was fired, the surgeon stated that there was an incomplete staple line.The surgeon cut off the staple line and re-inserted another anvil orvil and another stapler to complete the procedure.There was an unanticipated tissue loss due to the reported event.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.The visual inspection of the staple guide noted the instrument was fully applied.The orvil anvil used for all functional testing.Functionally, the device was applied over the appropriate test media producing acceptable results.The knife cut the test media cleanly and completely, and the pushers advanced.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.The investigation concluded there were no abnormalities that would have caused or contributed to the reported condition.If information is provided in the future, a supplemental report will be issued.
|
| |
|