• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY MITEK LLC US ARTHRO SUTURE MANIP *EA; SUTURE/NEEDLE PASSER, REUSABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US ARTHRO SUTURE MANIP *EA; SUTURE/NEEDLE PASSER, REUSABLE Back to Search Results
Catalog Number 214626
Device Problem Material Twisted/Bent (2981)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/09/2019
Event Type  malfunction  
Manufacturer Narrative
Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Udi: (b)(4).Incomplete the lot number is not currently available.As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported by the sales rep via phone that during a shoulder repair procedure, it was found that the middle of the metal shafts of the customer's knot pusher arthroscopic and suture manipulator grasper were slightly bent.The procedure was able to be completed with the devices with no patient harm or surgical delay to the case.The sales rep was not present for the case; therefore, could not provide any further information or lot numbers for the devices but stated that both devices were approximately two years and have seen heavy use in the field.The devices were discarded by the customer.There was no delay in the surgical procedure.There was a spare device available for use to complete the surgery.There was patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: investigation summary :the complaint device is not being returned, it was discarded by the customer, therefore unavailable for a physical evaluation.This complaint cannot be confirmed.One possible root cause of the reported problem could be fair wear and tear due to age and use.However without the return of the complaint device, and no further information provided we cannot determine a definitive root cause.Further, no lot numbers were supplied which precludes conducting a dhr review or a lot specific search in the complaints handling system.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
SUTURE/NEEDLE PASSER, REUSABLE
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
MDR Report Key9519820
MDR Text Key207435400
Report Number1221934-2019-60124
Device Sequence Number1
Product Code HCF
UDI-Device Identifier10886705004089
UDI-Public10886705004089
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 12/11/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number214626
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 12/11/2019
Initial Date FDA Received12/26/2019
Supplement Dates Manufacturer Received01/03/2020
Supplement Dates FDA Received01/06/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-