| Catalog Number |
TCM-445 |
| Medical Device Problem Code |
Patient-Device Incompatibility (2682)
|
| Health Effect - Clinical Code |
Capsular Contracture (1761)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Allergan did not submit this mdr within 30 days of becoming aware.Recent stimulated reporting related to 2011068-7/2/19-001-r has increased complaint and mdr volume.Allergan is implementing a plan to address the increased volumes.A review of the device history record has been initiated.If any new, changed or corrected information is noted, a supplemental medwatch will be submitted.The event of "capsular contracture" is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: "capsular contracture¿ ¿anxiety because of the implants." and "raynaud's syndrome.".
|
| |
|
Event or Problem Description
|
|
Patient reported right side ¿capsular contracture¿ baker grade unknown, ¿anxiety because of the implants." patient additionally reported "raynaud's syndrome " which is considered not device related.Device has been explanted.
|
| |
|
Search Alerts/Recalls
|