| Model Number |
68LP-225 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
10/31/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Allergan did not submit this mdr within 30 days of becoming aware.Recent stimulated reporting related to 2011068-7/2/19-001-r has increased complaint and mdr volume.Allergan is implementing a plan to address the increased volumes.A review of the device history record has been completed.No deviations or non-conformances noted.Device evaluation: visual analysis of the returned device identified: creases, wear abrasion, void >1 mm in non-textured valve-to shell-bond area and one linear opening.Leak test and microscopic analysis was performed which identified: one opening crease smooth.Fill inspection was performed and identified no blockage in the valve.Creases/folding of implant condition that was indicated in the complaint was observed, nevertheless it is considered non-related to the manufacturing process, since the device was received out of their primary packaging and is an explanted device.Based on the device analysis the final assessment is: one opening crease smooth assessed as fold flaw opening and creases/folding of implant condition was observed, nevertheless is not related to the manufacturing process.Reason for reoperation: deflation.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported deflation and "any creases".The device was explanted.This record is for the right side.
|
| |
|
Additional Manufacturer Narrative
|
|
Allergan did not submit this mdr within 30 days of becoming aware.Recent stimulated reporting related to 2011068-7/2/19-001-r has increased complaint and mdr volume.Allergan is implementing a plan to address the increased volumes.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported deflation and "any creases".The device was explanted.This record is for the right side.
|
| |
|
Search Alerts/Recalls
|