| Catalog Number |
140.588 |
| Medical Device Problem Code |
Failure to Osseointegrate (1863)
|
| Health Effect - Clinical Code |
Tissue Damage (2104)
|
| Date of Event |
10/21/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Considering the surgical methodology, it was seen that the dentist has used sequence of drills different than the one recommended for the implant installation.The investigation of the complaint was carried out considering the analysis of the information given by the dentist in the guarantee form, visual conference of the product returned by the dentist and the evaluation of relevant production records.
|
| |
|
Event or Problem Description
|
|
(b)(4) ¿ the dentist reported that almost 6 months after the dental implant was installed in patient¿s mouth, its non-osseointegration was observed.Moreover, 32n.Cm of primary stability was achieved, the patient presented insufficient bone quality/quantity (bone type iv) and bone graft was placed.
|
| |
|