• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. R3 20 DEG XLPE ACET LNR 28MM X 46MM; PROSTHESIS, HIP, SEMI-CONS, UNCEMENTED, METAL/POLY, POROUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. R3 20 DEG XLPE ACET LNR 28MM X 46MM; PROSTHESIS, HIP, SEMI-CONS, UNCEMENTED, METAL/POLY, POROUS Back to Search Results
Model Number 71334946
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Code Joint Dislocation (2374)
Date of Event 11/21/2018
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient underwent a closed reduction due to dislocation.
 
Additional Manufacturer Narrative
The devices, used in treatment, were not returned for evaluation and the reported event could not be confirmed.A clinical evaluation was conducted and confirms it cannot be determined to what extent the patient¿s history of spinal fusion, increased combined anteversion and soft tissue laxity had on her dislocation.The root cause of the dislocation post second revision cannot be determined and it cannot be concluded that the dislocation is associated with a mal-performance of the implant.The patient impact beyond the dislocation and reduction cannot be determined.A review of complaint history on the listed parts revealed no prior complaints for the listed batches with the same failure mode.A review of the device history records for the listed batch did not reveal any deviation from the standard manufacturing processes.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Some potential probable causes for this event could include improper surgical technique, device selection or patient anatomy.Based on this investigation, the need for corrective action is not indicated.Without the actual product involved, our investigation cannot proceed.If the devices or new information is received in the future, these complaints can be re-opened.No further actions are being taken at this time.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
R3 20 DEG XLPE ACET LNR 28MM X 46MM
Common Device Name
PROSTHESIS, HIP, SEMI-CONS, UNCEMENTED, METAL/POLY, POROUS
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key9520282
Report Number1020279-2019-04600
Device Sequence Number13602029
Product Code MBL
UDI-Device Identifier03596010598097
UDI-Public03596010598097
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K113848
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device No Information
Device Expiration Date10/17/2025
Device Model Number71334946
Device Catalogue Number71334946
Device Lot Number15KM13975
Initial Date Received by Manufacturer 12/09/2019
Supplement Date Received by Manufacturer04/09/2020
08/13/2021
Initial Report FDA Received Date12/26/2019
Supplement Report FDA Received Date04/14/2020
08/16/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-