Physio-control further evaluated the removed therapy pcb assembly and found the reported issue was due to transistor, designator q27.Pin 5 to 9 were shorted in diode, designator cr30, which was caused by burning from transistor, designator q26.The purpose of transistor, designator q27 is to aid in the turn off of cr30 which should lead to the device shut off after energy is delivered.The issue with transistor, designator q27 resulted in an initial shock being biphasic, but further shocks being monophasic and less energy when tested on a mock-up device, in either a defibrillation shock or during a user test.Device delivered shock, then turned off.The root cause of the reported issue was identified as transistor, designator q27.
|