It was reported that the device alarmed ail during a blood infusion programmed at a rate of 120ml/hr.The rn stated that she was unable to administer the blood due to frequent ail alarms, the lvp module was replaced 3 times as well as the tubing replaced 3 times , a new bag of blood ordered however the ail alarms continued.The rn stated that there was no visible air in the tubing when replaced the modules and sets.The icu/stepdown nurse eventually hung the blood to gravity to continue with the infusion.There was no report of patient impact however a delay in patient treatment.An incomplete event date of (b)(6) 2019 provided as customer stated that the event occurred 4 months ago.Although requested there was no additional event details provided by the customer.
|